The axial length of the eye and central corneal thickness is measured in this protocol using the Ultrasonic A scan/pachymeter DGH 4000B SBH IOL Computation module. (A-Scan pachymeter, Ultrasonic, Exton, PA) which requires direct contact with the cornea. Three measurements are taken on each eye for both axial length and central corneal thickness. Results of all measurements are then printed. The WG recommends that three readings be taken and recorded from each eye for both axial length and central corneal thickness (measurements in millimeters). The WG recommends that the central corneal thickness-pachymetry measurements be taken first followed by the axial length measurements, as presented in this protocol. This protocol for contact biometry uses the ultrasonic A scan/pachymeter DGH 4000B SBH IOL Computation module. Other instruments can be used to measure ocular biometry including axial length and central corneal thickness. If other instruments are used, the reproducibility of the measurements should be comparable to those acquired with this protocol. In addition, when other instruments are used to collect these measurements, the manufacturer and model of equipment should be recorded. These other devices may require some different steps than are described in this protocol. Investigators should follow the equipment manufacturer's instructions to ensure quality control.
Registration Status:
Qualified

Designations:

Designation:
Contact biometry proto
Tags:
Short Name
Designation:
PhenX - contact biometry protocol 110101
Tags:
Long Common Name

Designations:

Definition:
The axial length of the eye and central corneal thickness is measured in this protocol using the Ultrasonic A scan/pachymeter DGH 4000B SBH IOL Computation module. (A-Scan pachymeter, Ultrasonic, Exton, PA) which requires direct contact with the cornea. Three measurements are taken on each eye for both axial length and central corneal thickness. Results of all measurements are then printed. The WG recommends that three readings be taken and recorded from each eye for both axial length and central corneal thickness (measurements in millimeters). The WG recommends that the central corneal thickness-pachymetry measurements be taken first followed by the axial length measurements, as presented in this protocol. This protocol for contact biometry uses the ultrasonic A scan/pachymeter DGH 4000B SBH IOL Computation module. Other instruments can be used to measure ocular biometry including axial length and central corneal thickness. If other instruments are used, the reproducibility of the measurements should be comparable to those acquired with this protocol. In addition, when other instruments are used to collect these measurements, the manufacturer and model of equipment should be recorded. These other devices may require some different steps than are described in this protocol. Investigators should follow the equipment manufacturer's instructions to ensure quality control.
Tags:
Source: Regenstrief LOINC

Properties:

Key:
Related Names
Value:
PanPanelPANEL.PHENX
PanlPnlPoint in time
Random
Key:
Related Codes
Value:
Code SystemCodeCode TextCode Version
https://www.phenxtoolkit.org110101axial_length_using_contact_biometry
Key:
Fully-Specified Name
Value:
ComponentPropertyTimeSystemScaleMethod
PhenX - contact biometry protocol 110101-Pt^Patient-PhenX
Key:
Basic Attributes
Value:
ClassTypeFirst ReleasedLast UpdatedChange ReasonPanel Type
PANEL.PHENXClinicalVersion 2.36Version 2.66Updated the PhenX ID from "PhenX." to "PX" in Survey Question Source field to align with the variable identifier used in the PhenX Toolkit.; Added the PhenX protocol ID to the Component to clearly define the protocol version for which this panel is based upon.Panel

Identifiers:

Source:
NLM
Id:
Qkrl35zp3jg
Version:
1.0
Source:
PhenX
Id:
110101
Version:
Source:
LOINC
Id:
62679-6
Version:
2.69